Periodontal Probe Markings & Types Guide – UNC-15, Williams, Nabers & WHO Explained
Periodontal probes are the measuring instruments of periodontics — calibrated tips used to record pocket depths, attachment levels, bleeding on probing, and furcation involvement.
Read more about this guide +
But not all probes read the same way: marking patterns differ between designs, and choosing the wrong probe (or misreading its bands) leads to inconsistent charting. This guide explains the major periodontal probe types, their marking systems, and when to use each — with a comparison chart for quick reference. Note that German manufacturers — including Helmut Zepf, whose full probe range Deutsche Dental Technologien supplies — often catalog these patterns under different names: the UNC-15 is Helmut Zepf's North Carolina CNC, and the Marquis marking system appears as CPG 12; the chart links each pattern to its ZEPF counterpart.
What Do Periodontal Probe Markings Mean?
Periodontal probe markings are millimeter lines or color bands used to read pocket depth at the gum margin.
The UNC-15 is marked at every millimeter to 15 mm, the Williams at 1, 2, 3, 5, 7, 8, 9 and 10 mm (4 and 6 are skipped so the lines are easier to tell apart), the Marquis in alternating 3 mm color bands, and the WHO probe has a 0.5 mm ball tip with a black band from 3.5 to 5.5 mm.
Healthy pockets generally read 1–3 mm; 4 mm or more warrants closer evaluation.
Shop the probes in this guide · Helmut Zepf, made in Germany · ships from our US warehouse · 60-day returns
- UNC-15 North Carolina probe, single-ended (No. 24.453.23) — $39.99: every millimeter to 15
- Williams probe insert (No. 24.454.05) — $23.99: 1-2-3-5-7-8-9-10 mm, M 2.5 thread
- WHO probe CP 11.5, single-ended (No. 24.453.03) — $39.99: ball tip, 3.5-5.5-8.5-11.5 mm
- Marquis probe CPG 12, single-ended (No. 24.453.12) — $39.99: 3-6-9-12 mm
- Nabers furcation probe, double-ended (No. 24.453.15) — $49.99: 3-6-9-12 mm
Download the 4-page printable probe markings guide (PDF): all marking patterns drawn to one scale, probing technique and furcation grading.
Periodontal Probe Marking Systems: Comparison Chart
| Probe Type | Markings (mm) | Tip Design | Primary Use |
|---|---|---|---|
| UNC-15 (ZEPF: North Carolina CNC) | Every millimeter to 15; colored bands at 4–5, 9–10, and 14–15 mm | Straight, thin, round | Precise full-mouth charting; research standard |
| Williams | 1, 2, 3, 5, 7, 8, 9, 10 (4 and 6 omitted to reduce misreading) | Straight, round | General pocket depth measurement; the classic teaching probe |
| Michigan-O (ZEPF: Contrast Insert #1, 3/6/8/11) | 3, 6, 8 (ZEPF pattern adds 11) | Straight, fine, round | Quick screening with a slim tip |
| Marquis (ZEPF: CPG 12) | Alternating color bands every 3 mm (3, 6, 9, 12) | Straight, round | Rapid visual screening by band |
| WHO / CPITN (ZEPF: CP 11.5) | 0.5 mm ball tip; black band from 3.5–5.5 mm (some versions add 8.5 and 11.5) | Ball-ended | BPE / PSR screening; gentle on tissue |
| PSI Screening (ZEPF: CPN 22) | 2, 4, 6, 8, 10, 12 | Straight, round | Periodontal Screening Index recording at 2 mm intervals |
| Nabers | Color-coded, typically at 3, 6, 9, 12 | Curved, double-ended | Furcation involvement classification (Grades I–III) |
Probe Marking Patterns at a Glance
The diagrams below show each marking pattern drawn to the same millimetre scale, tip on the left, so the patterns can be compared directly. Reference bands are shown dark where the probe uses a black band and red where it uses a colour band; the numbers are the calibration points as they appear on the instrument.
The Same Patterns on the Instrument
Diagrams make the patterns easy to compare; the instruments show how they actually read chairside. Below are the Helmut Zepf probes and interchangeable inserts we stock, one for each pattern above, each photographed close on the graduated working end. Most are from the Contrast series (No. 24.451.xx), where the graduations are filled black so they stay legible against wet tissue — the clearest way to see how little the Michigan-O's three bands have in common with the UNC-15's fifteen. Select any image for the product.
1–15 mm · Contrast insert No. 24.451.06 Williams
1-2-3-5-7-8-9-10 mm · Insert No. 24.454.05 Michigan-O
3-6-8-11 mm · Contrast insert No. 24.451.00 Marquis / CPG 12
3-6-9-12 mm · Contrast insert No. 24.451.01 WHO / CPITN
3.5-5.5-8.5-11.5 mm · Contrast insert No. 24.451.03 PSI Screening / CPN 22
2-4-6-8-10-12 mm · Contrast insert No. 24.451.02 Nabers furcation probe
3-6-9-12 mm · Complete instrument No. 24.453.15
The Williams and Nabers patterns are shown with etched (rolled) graduations rather than contrast fill. Browse every pattern on the Periodontal Probe Insert Collection page, or the complete ZEPF Periodontal Probe Collection.
What Is a Marquis Probe?
A Marquis probe is a periodontal probe marked with alternating color bands every 3 mm — at 3, 6, 9, and 12 mm — rather than individual millimeter lines, so pocket depth is read by band at a glance.
Its straight, round tip makes it the pattern for rapid visual screening, while full-mouth charting is usually done with an every-millimeter probe such as the UNC-15.
Helmut Zepf catalogs the Marquis pattern as CPG 12: the single-ended Periodontal Probe CPG 12 (Marquis), No. 24.453.12 on a round handle, plus matching CPG 12 interchangeable inserts for handle systems that take replaceable tips.
What Is a Nabers Probe?
A Nabers probe is a curved, blunt-tipped periodontal probe used to find and measure furcation involvement — the bone loss between the roots of multi-rooted teeth. The curve is the whole point: it follows the root trunk around and into the furcation entrance, where a straight probe simply cannot go.
Markings are color-coded at 3, 6, 9 and 12 mm, so the horizontal depth the tip travels between the roots is read directly off the shank.
It is the one probe pattern with no substitute — furcation involvement cannot be staged reliably with a straight probe, and radiographs routinely understate it, particularly on maxillary molars where the roots overlap.
How Furcation Involvement Is Graded
Two classifications are in common use, and the Nabers reading feeds both. Hamp grades by horizontal probing depth: Degree I is less than 3 mm of horizontal penetration, Degree II is 3 mm or more but not through the tooth, and Degree III is through-and-through.
Glickman grades by what the probe and tissue show together: Grade I is incipient, with the flute palpable but no significant horizontal loss; Grade II is a cul-de-sac, bone lost on one aspect but not communicating; Grade III is through-and-through but still covered by gingiva; Grade IV is through-and-through and clinically visible.
Record the grade per furcation entrance, not per tooth — a maxillary molar has three.
Using a Nabers Probe
-
Know where the entrances are.
Mandibular molars furcate buccally and lingually. Maxillary molars have three: the buccal entrance mid-buccally, the mesial entrance approached from the palatal (it sits toward the palatal third), and the distal entrance reachable from either buccal or palatal.
-
Walk, don't poke.
Run the curved tip along the root surface at the cervical third until it drops into the concavity; light force, as with any probing.
-
Read horizontally.
The measurement that matters is how far the tip travels between the roots, not apically down a pocket.
-
Double-ended means paired.
The two curved ends mirror each other so one instrument reaches both the mesial and distal aspects without being rotated awkwardly in the mouth.
Nabers Probes from Helmut Zepf
Helmut Zepf catalogs the pattern under the Nabers name, in two versions: the Furcation Probe by Nabers, No. 24.453.15 — double-ended on a round handle, graduated 3-6-9-12 mm with rolled (laser-etched) markings that survive repeated autoclave cycles — and the TOUCHGRIP Furcation Curette/Probe by Nabers, No. 24.744.60HZU, designed by Dr. Hürzeler and Dr. Zuhr, which pairs the curved probe with a furcation curette on one double-ended instrument for periodontal and microsurgical work. Both are on the Nabers Furcation Probe page.
How to Read a Periodontal Probe
Insert the probe parallel to the long axis of the tooth with light pressure (approximately 10–25 g) until resistance is met at the base of the sulcus or pocket. Read the measurement at the gingival margin, rounding up to the nearest marking. Record six sites per tooth: mesio-buccal, buccal, disto-buccal, and the corresponding lingual sites.
Pocket depths of 1–3 mm are generally healthy; 4 mm and beyond warrants closer periodontal evaluation. Consistency matters more than any single reading — which is why practices standardize on one probe design across all operatories.
Periodontal Probe Measurements: What the Numbers Mean
A probe reading is the distance in millimeters from the gingival margin to the base of the sulcus or pocket. On its own it is the probing depth; combined with the position of the gingival margin, it gives the clinical attachment level, the measurement periodontal staging is based on.
| Probing Depth | What It Usually Indicates |
|---|---|
| 1–3 mm | Healthy sulcus, especially with no bleeding on probing |
| 4 mm | Early pocketing; note bleeding and recheck after debridement |
| 5–6 mm | Moderate pocketing; commonly treated with scaling and root planing |
| 7 mm or more | Deep pocketing; commonly evaluated for periodontal surgery or referral |
These ranges are a general guide. A diagnosis also depends on attachment loss, bleeding on probing, mobility and radiographs, not on pocket depth alone.
Probing Depth vs Clinical Attachment Level (CAL)
Probing depth is measured from the gingival margin, which moves: swelling pushes it up and recession pulls it down. Clinical attachment level is measured from the cemento-enamel junction (CEJ), which does not move, so it shows how much support the tooth has actually lost.
-
Gingival margin below the CEJ (recession):
CAL = probing depth + recession. A 4 mm pocket with 2 mm of recession is 6 mm of attachment loss.
-
Gingival margin above the CEJ (swelling or overgrowth):
CAL = probing depth − the distance from the CEJ to the margin. A 5 mm pocket with the margin 2 mm above the CEJ is 3 mm of attachment loss.
-
Gingival margin at the CEJ:
CAL equals the probing depth.
The 2017 AAP/EFP classification stages periodontitis by interdental CAL at the site of greatest loss: Stage I is 1–2 mm, Stage II is 3–4 mm, and Stages III and IV are 5 mm or more, with Stage IV adding complexity factors such as tooth loss. Because these thresholds are one millimeter apart, an every-millimeter probe such as the UNC-15 (North Carolina CNC) reads them more reliably than a banded screening probe.
Probing Technique, Step by Step
-
1Grasp and rest.
Hold the probe in a modified pen grasp with a stable finger rest close to the tooth being measured.
-
2Insert parallel.
Keep the tip against the root surface and the probe parallel to the long axis of the tooth; tilting it lengthens the measured path.
-
3Use light force.
About 10–25 g, enough to blanch a fingernail without discomfort.
-
4Walk the probe.
Move around the tooth in small 1 mm steps with the tip kept at the base of the sulcus, rather than lifting out and re-inserting.
-
5Record six sites per tooth.
Mesio-buccal, mid-buccal and disto-buccal, plus the three matching lingual or palatal sites, rounding up to the nearest marking.
-
6Note bleeding.
Record bleeding on probing that appears within 10–30 seconds, site by site; it is one of the most useful signs of active inflammation.
Common Probing Errors
-
Excessive force:
the most common error; in inflamed tissue it can inflate depths by 1 mm or more.
-
Incorrect angulation:
tilting away from the long axis overstates depth.
-
Missing the col area:
interproximal readings taken too far buccally miss the deepest point under the contact.
-
Calculus interference:
heavy deposits stop the tip short of the pocket base and understate depth.
-
Skipping furcations:
multi-rooted teeth charted with a straight probe only; without a Nabers probe at each furcation entrance, Grade II involvement is routinely missed.
-
Worn or bent tips:
a bent tip adds the same systematic error to every site (see Care and Sterilization below).
Choosing the Right Probe
Helmut Zepf Periodontal Probes
Our ZEPF Periodontal Probe Collection covers the marking systems above in surgical stainless steel, with laser-etched calibrations that survive repeated autoclave cycles without fading — a common failure point on printed-band probes. Several handle systems accept interchangeable probe inserts, letting you replace a worn tip without replacing the instrument.
Browse the full periodontal probe collection or the complete ZEPF Hygiene Instrument Collection for the surrounding armamentarium, including Gracey curettes and sickle scalers. Exploring the maxillary antrum rather than the sulcus? See the Antrum & Surgical Probe Collection.
Care and Sterilization
Probes are precision-calibrated instruments: avoid ultrasonic baskets that allow tip-to-tip contact, inspect calibration bands during reprocessing, and retire probes with bent or worn tips — a 0.5 mm bend at the tip is a 0.5 mm charting error on every patient. Full reprocessing parameters are in our Instrument Reprocessing & Sterilization Guide. For terminology, see the Hygiene Instrument Glossary.
Frequently Asked Questions
What do the markings on a perio probe mean?
The markings are millimeter calibrations used to measure pocket depth from the gingival margin to the base of the sulcus. Different probe designs mark different intervals — the UNC-15 marks every millimeter, the Williams skips 4 and 6, and the Marquis uses 3 mm color bands.
What is the difference between a UNC-15 and a Williams probe?
The UNC-15 is marked at every millimeter to 15 mm, with colored bands at 4–5, 9–10, and 14–15 mm for orientation. The Williams probe marks 1, 2, 3, 5, 7, 8, 9, and 10, deliberately omitting 4 and 6 so adjacent lines are easier to distinguish.
Is the North Carolina CNC probe the same as a UNC-15?
Yes. UNC stands for University of North Carolina — German manufacturers including Helmut Zepf catalog the same 1–15 mm every-millimeter pattern as "North Carolina" or "CNC."
What is a Nabers probe used for?
The Nabers probe is a curved, color-coded probe designed to enter and measure furcation areas between the roots of multi-rooted teeth, used to classify furcation involvement from Grade I to Grade III.
What is the difference between a Nabers probe and a regular periodontal probe?
Shape and direction of measurement. A straight probe measures vertically, from the gingival margin down to the base of the pocket. The Nabers probe is curved so it can reach around the root trunk into a furcation entrance, and the reading that matters is horizontal — how far the tip travels between the roots, graduated at 3, 6, 9 and 12 mm.
How do you grade a furcation with a Nabers probe?
By horizontal penetration at each furcation entrance. Under the Hamp classification, less than 3 mm is Degree I, 3 mm or more without passing through is Degree II, and through-and-through is Degree III. Glickman's system grades I to IV, with Grade III through-and-through but covered by gingiva and Grade IV clinically visible. Grade each entrance separately — a maxillary molar has three.
How much pressure should be used when probing?
Approximately 10–25 grams of force — commonly described as the pressure that blanches a fingernail without causing discomfort. Excessive force penetrates the junctional epithelium and overstates pocket depth.
What is a normal periodontal probing depth?
1–3 mm is generally consistent with health. Readings of 4 mm or more, especially with bleeding on probing, warrant further periodontal evaluation.
How do you calculate clinical attachment level from probe measurements?
Add the recession to the probing depth when the gingival margin sits below the cemento-enamel junction; subtract the distance from the CEJ to the margin when the margin sits above it. When the margin is at the CEJ, attachment level equals probing depth.
What is the difference between a perio probe and an explorer?
A probe has a blunt, calibrated tip for measuring depth in millimeters; an explorer has a sharp, flexible tip for detecting calculus, caries and margin defects by touch. Many diagnostic instruments combine one of each on a double-ended handle.
Can one probe replace all the others?
No single marking pattern covers every task well. Most practices keep a UNC-15 for comprehensive charting, a color-banded probe such as the Marquis (CPG 12) for quick screening, and a Nabers probe for furcation checks.
Are Helmut Zepf periodontal probes under warranty?
Yes. Every probe in this guide is manufactured by Helmut Zepf in Seitingen-Oberflacht, Germany, and sold in the U.S. by Deutsche Dental Technologien, which backs the line with its standard warranty and 60-day return policy — see our About Us page for details.
4-page printable quick reference — all seven marking patterns drawn to one scale, probing technique, CAL, and furcation grading.